Event Description
A customer reported to olympus, the evis exera iii xenon light source image would go in and out when wiggling the scope.The event occurred during preparation for use.There was no report of patient harm.The device was returned for evaluation.During the device evaluation, the following reportable malfunctions were identified: b30 (scope communication) error was displayed, and illumination had excessive thermal paste.
 
Manufacturer Narrative
The subject device was returned to olympus for evaluation.During inspection and testing, the allegation was confirmed due to a damaged scope socket.When the connected scope was slightly moved or bumped, the image is lost and a b30 (scope communication) error was displayed.In addition to evaluation in b5, the front panel was chipped.The lamp¿s life had over 500 hours.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 7 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the damaged/faulty scope socket could not be identified.Olympus will continue to monitor field performance for this device.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the device evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
Additional information was provided regarding this event.The intended procedures were diagnostic esophagogastroduodenoscopy (egd) and colonoscopy.The procedures were delayed by less than 5 minutes and did not require additional treatment.The procedures were completed using the subject device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17097794
MDR Text Key317744598
Report Number3002808148-2023-05775
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/05/2023
Initial Date FDA Received06/09/2023
Supplement Dates Manufacturer Received06/12/2023
07/07/2023
Supplement Dates FDA Received06/20/2023
08/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-