This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d9, g3, g6, h1, h2, h3, h4, h6, h10 reported event was confirmed by review of x-rays provided.Visual evaluation of the returned device shows signs of implantation with scratches, nicked, and foreign material on the distal surface.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.X-ray review indicates there is femoral notching is present.Minimal lucency also seen along the cement hardware interface of the femoral component on the lateral film.Moderate joint effusion.Ossific density within the posterior joint space could represent fabella versus loose body.Overall fit and alignment of the implants is appropriate.Radiolucency along the hardware indicating possible loosening of the femoral and tibial components.Suggestion of minimal cement on the ap film involving the medial aspect of the tibial component.More prominent appearance of cement within the central aspect of the distal femur.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|