An instrument check was performed and all tests passed.Upon review of the alarm trace, there were no sample-related alarms around the time the patient sample was pipetted.The alarm trace showed multiple sample foam detection alarms, sample clot, and sample short alarms occurring between may 2023 and 05-jun-2023.The investigation could not identify a product problem.The cause of the event could not be determined.
|
The initial reporter stated they received discrepant results for one patient sample tested with elecsys troponin t hs on cobas e 801 analytical unit serial number (b)(6).The sample initially resulted in a troponin t value of 10.5 ng/l and this value was reported outside of the laboratory.The doctor questioned the result as it was inconsistent with the patient's previous result.The doctor requested a repeat of the sample.The sample was repeated, resulting in a troponin t value of 4025 ng/l.The sample was also repeated on a second e 801 analyzer, resulting in a troponin t value of 3996 ng/l.
|