• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA LRS PLATELET PLASMA RBC AUTOPAS SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA LRS PLATELET PLASMA RBC AUTOPAS SET Back to Search Results
Model Number 82410
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problem Hypersensitivity/Allergic reaction (1907)
Event Date 01/20/2023
Event Type  Injury  
Event Description
The customer reported an adverse reaction from trima platelets.During the procedure the patient presented with erythematous papules on the face and scalp and was given iv chlorpheniramine and iv hydrocortisone.The patient is reported as stable.The collection set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
Investigation is in process, a follow-up report will be provided.
 
Event Description
The customer reported an adverse reaction from trima platelets.During the procedure the patient presented with erythematous papules on the face and scalp and was given iv chlorpheniramine and iv hydrocortisone.The patient is reported as stable.The collection set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Investigation: a disposable complaint history search was performed for this lot and found no reports for similar issues on this lot.According to the aabb circular of information for the use of human blood components (revised 2017), allergic reactions frequently occur (ie, 1-3% of plasma-containing components) as mild or self-limiting urticaria or wheezing that usually respond to antihistamines.More severe manifestations, including respiratory and cardiovascular symptoms, are more consistent with anaphylactoid/anaphylactic reactions and may require more aggressive therapy.No laboratory procedures are available to predict these reactions.Anaphylactic reactions, characterized by hypotension, tachycardia, nausea, vomiting and/or diarrhea, abdominal pain, severe dyspnea, pulmonary and/or laryngeal edema, and bronchospasm and/or laryngospasm, are rare (<10/100,000 transfused units) but dangerous complications requiring immediate treatment with epinephrine.While these reactions have been reported in iga-deficient patients with anti-iga antibodies and patients with haptoglobin deficiency, most reactions are idiosyncratic and not associated with a specific serum protein deficiency, polymorphism, or identifiable cause.A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.Investigation is in process.A follow-up report will be provided.
 
Event Description
The customer reported an adverse reaction from trima platelets.During the procedure the patient presented with erythematous papules on the face and scalp and was given iv chlorpheniramine and iv hydrocortisone.The patient is reported as stable.The collection set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Investigation: a disposable complaint history search was performed for this lot and found no reports for similar issues on this lot.According to the aabb circular of information for the use of human blood components (revised 2017), allergic reactions frequently occur (ie, 1-3% of plasma-containing components) as mild or self-limiting urticaria or wheezing that usually respond to antihistamines.More severe manifestations, including respiratory and cardiovascular symptoms, are more consistent with anaphylactoid/anaphylactic reactions and may require more aggressive therapy.No laboratory procedures are available to predict these reactions.Anaphylactic reactions, characterized by hypotension, tachycardia, nausea, vomiting and/or diarrhea, abdominal pain, severe dyspnea, pulmonary and/or laryngeal edema, and bronchospasm and/or laryngospasm, are rare (<10/100,000 transfused units) but dangerous complications requiring immediate treatment with epinephrine.While these reactions have been reported in iga-deficient patients with anti-iga antibodies and patients with haptoglobin deficiency, most reactions are idiosyncratic and not associated with a specific serum protein deficiency, polymorphism, or identifiable cause.A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.Root cause: a root cause assessment was performed for the allergic reactions.Based on the available information a definitive root cause could not be determined but it is likely due to one or a combination of the possible causes listed below: * hypersensitivity to the ethylene oxide used to sterilize the disposable set.* hypersensitivity to the components inside the disposable set.* anaphylactic reactions have been reported in iga-deficient patients with anti-iga antibodies and patients with haptoglobin deficiency, most reactions are idiosyncratic and not associated with a specific serum protein deficiency, polymorphism, or identifiable cause.* immediate hypersensitivity reaction via production of ige antibodies in response to allergens in the transfused product.* passive sensitization via passive transfer of donor allergens or donor antibodies (e.G.Allergen-specific ige); transient allergic sensitivity to allergens.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA LRS PLATELET PLASMA RBC AUTOPAS SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key17101450
MDR Text Key316931771
Report Number1722028-2023-00207
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583824102
UDI-Public05020583824102
Combination Product (y/n)N
Reporter Country CodeCI
PMA/PMN Number
BK190332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 06/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number82410
Device Catalogue Number82410
Device Lot Number2211022342
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/15/2023
Initial Date FDA Received06/09/2023
Supplement Dates Manufacturer Received06/19/2023
07/04/2023
Supplement Dates FDA Received06/30/2023
07/11/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age43 YR
Patient SexMale
Patient Weight75 KG
-
-