Model Number 7122Q/65
Device Problems High impedance (1291); Retraction Problem (1536); Capturing Problem (2891)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/23/2023
Event Type  Injury  
Event Description
It was reported that the patient presented for follow -up.A device interrogation revealed high pacing impedance, elevated capture threshold exhibited by the right ventricular lead.The patient was scheduled for replacement on (b)(6) 2023.During the procedure it was observed that the helix would not retract.The lead was subsequently capped and replaced.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
DEFIBRILLATION LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17102085
MDR Text Key316932423
Report Number2017865-2023-22851
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734503341
UDI-Public05414734503341
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2018
Device Model Number7122Q/65
Device Catalogue Number7122Q-65
Device Lot NumberA000034729
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/23/2023
Initial Date FDA Received06/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/26/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUADRA ASSURA; QUARTET; TENDRIL
Patient Outcome(s) Required Intervention;
-
-