ARROW INTERNATIONAL LLC ARROW CVC SET: 16 GA X 8" (20 CM); CATHETER PERCUTANEOUS Back to Search Results
Catalog Number ES-04301
Device Problems Migration or Expulsion of Device (1395); Off-Label Use (1494)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/30/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported "the drainage tube was inserted in the emergency room on (b)(6) 2022.The patient accidentally pulled out the right thoracic drainage tube while turning over in a lateral position on the evening of (b)(6) 2022.The patient has not informed the nurse yet.Around 2am, the patient requested the nurse to disinfect and change the dressing of the drainage tube, but the nurse found that the patient's drainage tube had come out and was fixed with a seamless thread.After the nurse disinfected the wound and changed the dressing, they informed the on duty doctor and the doctor checked the patient by the bedside." the patient's condition was reported as fine.
 
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event Description
It was reported "the drainage tube was inserted in the emergency room on (b)(6) 2022.The patient accidentally pulled out the right thoracic drainage tube while turning over in a lateral position on the evening of (b)(6) 2022.The patient has not informed the nurse yet.Around 2am, the patient requested the nurse to disinfect and change the dressing of the drainage tube, but the nurse found that the patient's drainage tube had come out and was fixed with a seamless thread.After the nurse disinfected the wound and changed the dressing, they informed the on duty doctor and the doctor checked the patient by the bedside." the patient's condition was reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 16 GA X 8" (20 CM)
Type of Device
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key17105397
MDR Text Key317772180
Report Number3006425876-2023-00560
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K862056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberES-04301
Device Lot Number71F21L0047
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/15/2023
Initial Date FDA Received06/12/2023
Supplement Dates Manufacturer Received06/30/2023
Supplement Dates FDA Received07/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NOT REPORTED; NOT REPORTED
-
-