RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Failure to Read Input Signal (1581)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/16/2023 |
Event Type
malfunction
|
Event Description
|
It was reported, while outside of clinical use, there was a spi bus failure identified in the event log.Data acquisition board to be replaced.No reports of harm or injury to patient or user.Investigation is ongoing.
|
|
Manufacturer Narrative
|
E1: phone number: institution: (b)(6).
|
|
Manufacturer Narrative
|
H10: the pse provided analysis findings to the customer and recommended replacement of the data acquisition (da) power management (pm) printed circuit board assembly (pcba).Once customer approval was received, the pse replaced the pm pcba.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.Components have not been received for evaluation at this time.If a part is received and evaluated, an additional report will be submitted.
|
|