Model Number V60
Device Problem Failure to Read Input Signal (1581)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/16/2023
Event Type  malfunction  
Event Description
It was reported, while outside of clinical use, there was a spi bus failure identified in the event log.Data acquisition board to be replaced.No reports of harm or injury to patient or user.Investigation is ongoing.
 
Manufacturer Narrative
E1: phone number: institution: (b)(6).
 
Manufacturer Narrative
H10: the pse provided analysis findings to the customer and recommended replacement of the data acquisition (da) power management (pm) printed circuit board assembly (pcba).Once customer approval was received, the pse replaced the pm pcba.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.Components have not been received for evaluation at this time.If a part is received and evaluated, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key17107057
MDR Text Key317579727
Report Number2518422-2023-13312
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838009851
UDI-Public00884838009851
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV60
Device Catalogue Number1053614
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/05/2023
Initial Date FDA Received06/12/2023
Supplement Dates Manufacturer Received07/06/2023
Supplement Dates FDA Received07/14/2023
Date Device Manufactured12/03/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-