This mdr is being submitted as part of a retrospective review and remediation effort based on enhancements made to the company¿s mdr and complaint handling processes.Capas have been opened to manage the actions that are being taken to remediate this issue and ensure any required mdr reporting is completed.The device was returned to olympus for evaluation where service confirmed the customer's complaint.Additionally, there was pump unit failure, the chassis had poor appearance, the rear panel had poor appearance, the output socket failed, there were image issues, and a non-olympus lamp was found.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the air pump failed.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
|