BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 367986
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/17/2023
Event Type  malfunction  
Manufacturer Narrative
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported when using the bd vacutainer® sst¿ blood collection tube the cap was broken.The following information was provided by the initial reporter.The customer stated: "the tube cap was found to be damaged.".
 
Manufacturer Narrative
H.6.Investigation summary: "no photos or customer samples were provided for the investigation.30 retain samples were subjected to a visual inspection for cracked hemogard shield.No cracked hemogard shields were found.Therefore, bd is unable to confirm the customers reported defect based on retain testing.The device history record was reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.This complaint is unable to be confirmed.".
 
Event Description
It was reported when using the bd vacutainer® sst¿ blood collection tube the cap was broken.The following information was provided by the initial reporter.The customer stated: "the tube cap was found to be damaged.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17113501
MDR Text Key317159453
Report Number1024879-2023-00367
Device Sequence Number1
Product Code JKA
UDI-Device Identifier50382903679862
UDI-Public50382903679862
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
BK050036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/30/2023
Device Model Number367986
Device Catalogue Number367986
Device Lot Number2179947
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/17/2023
Initial Date FDA Received06/12/2023
Supplement Dates Manufacturer Received06/14/2023
Supplement Dates FDA Received06/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-