• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND IN.PACT AV ACCESS; DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND IN.PACT AV ACCESS; DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLAS Back to Search Results
Catalog Number IAV12004008P
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hemorrhage/Bleeding (1888); Pain (1994); Stenosis (2263); Swelling/ Edema (4577)
Event Date 11/10/2022
Event Type  Injury  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
During the index procedure one in.Pact av access balloon catheter was used to treat the swing point and central vein of the left arm.Approximately 12 months post procedure the patient suffered from left central vein restenosis and subclavian vein restenosis.Prolonged bleeding, left arm swelling and achiness and occasional difficulty in cannulation were reported.The event was treated with a percutaneous intervention.The patient recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IN.PACT AV ACCESS
Type of Device
DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLAS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17115828
MDR Text Key317116267
Report Number9612164-2023-02541
Device Sequence Number1
Product Code ONU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P190008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/15/2023
Device Catalogue NumberIAV12004008P
Device Lot Number0010095591
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/17/2023
Initial Date FDA Received06/13/2023
Date Device Manufactured01/16/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age45 YR
Patient SexMale
Patient Weight91 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-