GENZYME CORPORATION(RIDGEFIELD) SYNVISC; MOZ Back to Search Results
Lot Number CRSP014A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Peripheral Edema (4578); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 01/18/2023
Event Type  Injury  
Event Description
Yesterday calf was 2x the size of other; gone downsome since last night [calf swelling] left knee is very swollen [injection site joint swelling] knee is a little more warm to the touch [injection site joint warmth] case narrative: initial information received on 07-mar-2023 regarding an unsolicited valid serious case received from a pharmacist via health authorities of united states under reference mw5115146.This case is linked to duplicate case (b)(4).This case involves an unknown age male patient whose calf was 2x the size of other yesterday; gone down some since last night, left knee was very swollen and knee was a little more warm to the touch after the use of medical device hylan g-f 20, sodium hyaluronate (synvisc).The patient's past medical history, medical treatment(s), concomitant(s) and family history were not provided.On (b)(6) 2023, the patient received second injection of hylan g-f 20, sodium hyaluronate injection in his left knee via intra-articular route (strength:16 mg/2 ml) (lot: crsp014a and expiry date: 31-aug-2025) (unknown indication, dose, frequency).Patient had his 2nd of 3 synvisc.Patient stated his left knee was very swollen (joint swelling) (latency: same day).The swelling started about 3pm yesterday afternoon.He stated yesterday his calf was 2x the size of his other, that has gone down some since last night (peripheral swelling) (latency: same day).The knee was a little more warm to the touch (joint warmth) (latency: same day), no redness.He did not have a fever.He had been using ice.Patient could not get here for an appointment before 4:30pm.It was unknown if the patient experienced any additional symptoms/events.It was unknown if there were lab data/results available.Action taken: unknown for all events.Corrective treatment: patient had been using ice for all events.Outcome: recovering for peripheral swelling and unknown for other events.Seriousness criteria: medically significant for all events.A ptc (product technical complaint) was initiated on 08-mar-2023, for synvisc (lot: crsp014a and expiry date: 31-aug-2025) with global ptc number (b)(4).The sample was not available.The ptc stated that based on the complaint from intake team, there was no quality related defect that would attribute to a death or serious injury.Sanofi global pharmacovigilance and epidemiology continuously monitored adverse event reported with or without lot numbers, and assesses possible associations with their corresponding product lot, as part of routine safety surveillance effort to detect safety signals.The defect class has been updated to ii.(tj (b)(6) 2023).Batch # crsp014a, synvisc was manufactured on 20sep2022 with expiration date of 31aug2025 yielding (b)(4) kits.The incoming component inspection, packaging, and quality control documentation for this lot was reviewed.The investigation showed the product met specification at the time of release.It was the requirement to review all finished batch records for specification conformance prior to release.Any out of specification result is identified and mitigated through the nonconforming material or product process.Furthermore, no associated non-conformances were noted for the issue defect at time of release.Customer sample was not available.Root cause could not be determined.Trend analysis: there were a total of 5 complaints for mother lot crsp0104 and sub-batches.100292748 crsp014a plunger missing100302463 crsp014a leakage at injection site/ adverse event / syringe issue nos100310438 crsp014a adverse event 100310443 crsp014a adverse event100310469 crsp014a.Adverse event there was no quality related defect that would attribute to a death or serious injury.Sanofi global pharmacovigilance and epidemiology continuously monitors adverse event reports with or without lot numbers, and assesses possible associations with their corresponding product lot, as part of routine safety surveillance effort to detect safety signals.This review has not indicated any safety issue.Based on investigation and trend analysis, no capa (corrective and preventive action) required.Sanofi will continue to adverse events.Trend analysis will be performed on a periodic basis "product event handling" to determine if a capa was required.The final investigation was completed on 20-apr-2023 with summarized conclusion as no assessment possible.Upon internal review, case (b)(4) was identified to be duplicate of (b)(4).Hence, all the information from the case (b)(4) (to be deleted) has been merged in case (b)(4).Case (b)(4) received with clock start date of 10-mar-2023 would be deleted.Ptc details, strength, batch number and expiry date added.Text amended accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNVISC
Type of Device
MOZ
Manufacturer (Section D)
GENZYME CORPORATION(RIDGEFIELD)
1125 pleasantview terrace
ridgefield 07657
Manufacturer (Section G)
GENZYME CORPORATION(RIDGEFIELD)
1125 pleasantview terrace
ridgefield 07657
Manufacturer Contact
heather schiappacasse
55 corporate drive, ms 55b-220
a
bridgewater 08807
MDR Report Key17116081
MDR Text Key317061980
Report Number2246315-2023-00065
Device Sequence Number1
Product Code MOZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 06/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Lot NumberCRSP014A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient SexMale
-
-