Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/22/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).H3- the product has been returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that a drill pin and screw inserter fractured.There is no additional information available.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.Visual examination of the provided pictures identified the unit as broken.Additionally, the features of the fracture surface visually align with section 5.8.1 in zrm_wa_0545_21 in which an overload fracture failure mode was identified.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no related deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is received which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL PIN AND SCREW INSERTER
Type of Device
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17118908
MDR Text Key317596946
Report Number0001822565-2023-01569
Device Sequence Number1
Product Code HWB
UDI-Device Identifier00889024201637
UDI-Public(01)00889024201637(10)65245978
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00590102100
Device Lot Number65245978
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/22/2023
Initial Date FDA Received06/13/2023
Supplement Dates Manufacturer Received07/31/2023
Supplement Dates FDA Received08/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/04/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-