MAUDE Adverse Event Report: MEDTRONIC MEXICO CUSTOM TUBING PACK; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO CUSTOM TUBING PACK; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO Back to Search Results
Model Number BB8G55R15
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/16/2020
Event Type  malfunction  
Event Description
Medtronic received information that prior to use of a custom tubing pack, during initial set up the customer noticed during priming that the stopcock was broken inside of the arterial line when it was screwed into place.The device was replaced to complete the procedure.There was no patient involvement so no adverse effect occurred.Additional information: there was no damage reported to the outer box.There was no other damage reported to any other component.
 
Manufacturer Narrative
Device evaluation summary: visual inspection shows the stopcock luer stem is broken off inside the mating luer connection on the arterial filter.Reason for return was confirmed.Conclusion: complaint is confirmed for damaged component.Analysis found that the stopcock luer stem was broken off inside the mating luer connection on the arterial filter; there is a possibility that this was caused by a defect produced by a supplier process failure.After evaluation the cause of this complaint could not be determined; though, medtronic has notified the component supplier to be aware of this condition.Review of the device history record found no abnormalities during manufacturing that would cause or contribute to the reported event.Assessment against the medtronic risk management file pfmeca document fmea0002-06 indicates that the current risk zone does not exceed the risk zone predicted in the product pfmeca.There were no adverse patient effects as a result of this incidence.Medtronic will continue to monitor for future occurrences and trends per trend analysis procedure.This regulatory report is being submitted as part of a retrospective review and remediation per d00953163 as part of a capa action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM TUBING PACK
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17125405
MDR Text Key317296678
Report Number9612164-2023-02588
Device Sequence Number1
Product Code DWF
UDI-Device Identifier00763000373955
UDI-Public00763000373955
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K171979
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/29/2022
Device Model NumberBB8G55R15
Device Catalogue NumberBB8G55R15
Device Lot Number220999856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/16/2020
Initial Date FDA Received06/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-