BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL Back to Search Results
Model Number M00510890
Device Problems Material Deformation (2976); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2023
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that a trapezoid rx was to be used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2023.During preparation, a trapezoid rx basket was unpacked, however, it was noted that the four wires at the end of the basket were twisted together and could not be deployed normally, and after several attempts, still to no avail.The procedure was completed with another trapezoid rx basket.There were no patient complications as a result of this event.The investigation results revealed that the side car-rx was pushed back; therefore, this is now an mdr reportable event.Please see block h10 for full investigation details.
 
Manufacturer Narrative
Block h6: imdrf device code a0406 captures the reportable investigation finding of side car rx pushback.Block h10: the returned trapezoid basket was analyzed, and a dimensional inspection noted that the side car rx was pushed back approximately 7.0mm, which is out of specifications.In addition, the wires at the end of the returned basket were in good conditions.The reported event was not confirmed.Based on all available information, the wires of the basket were not twisted and did not have any damage, therefore, cause code no problem detected was added.The investigation also concluded that the side car rx push back could have occurred due to procedural factors such as the device advancing though the scope that had led to this event.Therefore, cause code adverse event related to procedure was added.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17131741
MDR Text Key317844099
Report Number3005099803-2023-02933
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296409
UDI-Public08714729296409
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/27/2023
Device Model NumberM00510890
Device Catalogue Number1089
Device Lot Number0029839706
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/26/2023
Initial Date FDA Received06/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
Patient Weight41 KG
-
-