MEDTRONIC IRELAND RESOLUTE ONYX RX; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number RONYX30034UX
Device Problems Entrapment of Device (1212); Difficult to Remove (1528); Deformation Due to Compressive Stress (2889); Difficult to Advance (2920); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/03/2023
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
An attempt was made to use one resolute onyx rx coronary drug-eluting stent to treat a lesion.It was reported that the stent came off the balloon catheter in the guide catheter.The guide catheter was removed with the stent inside.It was also detailed that there were two wires twisted in the guide catheter.The procedure was completed with another medtronic stent.No injury to the patient was reported.
 
Manufacturer Narrative
Product analysis: the device was returned for analysis.Kinks were evident to the hypo-tube.The stent was not present on the balloon and did not return for analysis.A kink was evident to the distal shaft.Bunching was evident the distal shaft immediately proximal to the proximal balloon bond.The balloon folds were partially expanded.Crimp impressions were visible on the exposed balloon surface.The inner lumen patency was verified with a 0.015-inch mandrel.No deformation evident to the distal tip.No other damage evident to the remainder of the device.Correction: the device did not pass through a previously deployed stent or cell of a stent.Annex a code.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: the device was inspected with no issues noted.The device was prepped per ifu with no issues noted.Resistance was not noted during the delivery of the device.Excessive force was not used.It was stated that there was a slight difficulty advancing and retracting the device due to the wire twist.The guidewires were flushed with no kinks or issues noted.The replacement stent was delivered with no issues.The patient is reported to be fine.Initial reporter name and phone number provided.Correction: facility address.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: resistance was noted during the withdrawal of the device.The device was kinked but not re-straightened during use.Everything was taken out.A non mdt 6f 3.5 guide catheter was used during the procedure.Non mdt guidewires were used during the index procedure.There were no difficulties noted when loading the device onto the guidewire.The guidewire was used successfully with other devices prior to the difficulties occurring.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17134606
MDR Text Key317320011
Report Number9612164-2023-02614
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier00643169557031
UDI-Public00643169557031
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 09/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/05/2023
Device Model NumberRONYX30034UX
Device Catalogue NumberRONYX30034UX
Device Lot Number0225122089
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/06/2023
Initial Date FDA Received06/15/2023
Supplement Dates Manufacturer Received07/14/2023
08/17/2023
08/31/2023
Supplement Dates FDA Received07/27/2023
08/23/2023
09/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-