ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ISOFLEX OPTIM LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1948/58
Device Problems Failure to Capture (1081); Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/27/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.It was noted that the right ventricular lead failed to capture and failed to sense.The lead was not used and replaced during procedure.The patient was stable.
 
Manufacturer Narrative
The reported events of failure to capture and failure to sense were not confirmed.As received, a complete lead was returned in one piece.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray examinations of the lead found did not find any anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOFLEX OPTIM LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17140531
MDR Text Key317720381
Report Number2017865-2023-23379
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734502351
UDI-Public05414734502351
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P960030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1948/58
Device Catalogue Number1948-58
Device Lot NumberP000161869
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/27/2023
Initial Date FDA Received06/15/2023
Supplement Dates Manufacturer Received07/21/2023
Supplement Dates FDA Received07/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-