Based on the claim against the product by the customer noting fenestrated bipolar forceps stuck in trocar, an investigation was completed to determine the cause of this reported event.Intuitive surgical, inc.(isi) received a da vinci product to perform failure analysis.The fenestrated bipolar forceps was analyzed and the reported failure was confirmed.The instrument was found to have the main tube broken.Pieces measuring approximately 0.044¿ x 0.331¿ were not returned with the instrument.However, the instrument could not be placed on an in-house system for any functional testing due to the inability to feed through the cannula.The complaint regarding fenestrated bipolar forceps stuck in the trocar was confirmed by failure analysis.
|
It was reported that during a da vinci-assisted surgical procedure, that the fenestrated bipolar forceps joint would not pivot to come out of the robotic trocar.The procedure was completed with no reported patient injury.Intuitive surgical inc.(isi) made multiple follow-up attempts to obtain additional information.However, no further details have been received as of the date of this report.
|