Model Number 623-00-40G
Device Problem Unstable (1667)
Patient Problem Joint Laxity (4526)
Event Date 05/24/2023
Event Type  Injury  
Event Description
As reported: "dr.Performed a right hip revision due to instability.No more information is available per doctor.".
 
Manufacturer Narrative
An event regarding instability involving a trident liner was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to instability.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.The following devices were also listed in this report: cat# 6051-0935s; secur-fit max 132 hip stem #9; lot# mhne4r; cat# 19-0000t; uni adaptor sleeve c taper ti; lot# 33137804; cat# 6519-1-040; delta un.Fem hd 40mm r40 16/18; lot# 33808902; cat# 6704-0-520; d-m 2.0mm beaded cable set vit; lot# 33932103; cat# 6704-0-520; d-m 2.0mm beaded cable set vit; lot# 31319005.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.H3 other text : not returned to the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 0 DEG INSERT 40MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17141218
MDR Text Key317302587
Report Number0002249697-2023-00680
Device Sequence Number1
Product Code LPH
UDI-Device Identifier07613327025361
UDI-Public07613327025361
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061434
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 06/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2015
Device Model Number623-00-40G
Device Catalogue Number623-00-40G
Device Lot NumberMJKHRL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/24/2023
Initial Date FDA Received06/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/18/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age52 YR
Patient SexFemale
Patient Weight115 KG
-
-