MEDTRONIC SINGAPORE OPERATIONS AZURE XT DR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE
|
Back to Search Results |
|
Model Number W1DR01 |
Device Problems
Intermittent Capture (1080); Failure to Capture (1081); High impedance (1291); Mechanical Problem (1384); Over-Sensing (1438); Device Sensing Problem (2917)
|
Patient Problems
Bradycardia (1751); Fatigue (1849); Dizziness (2194)
|
Event Date 05/09/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient experienced low heart rate causing dizziness when standing and general fatigue.The right atrial (ra) lead was dislodged as confirmed on x-ray as was sometimes capturing the ventricle.The ra lead was explanted and replaced.It was reported that the right ventricular (rv) lead exhibited high undefined impedance with high and intermittent rv capture however when the rv lead was tested through analyzer everything was completely normal.The rv lead was tried in the atrial port and there was over sensing with no capture.A set screw/ header issue was suspected and when the new ra lead was plugged into the atrial port of the implantable pulse generator (ipg) no capture and inappropriate sensing were noted but was working fine through the analyzer.The ipg was explanted and replaced.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|