U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND VELYS SATELLITE STATION; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND VELYS SATELLITE STATION; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 4515-70-101
Device Problem Positioning Failure (1158)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/30/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Event Description
It was reported that the robotic assisted satellite station device, while being used with a robotic assisted base station device, the robot would not hold position.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was evaluated and the excessive movement on axis one of the robot was confirmed.The evaluation determined that the shaft axis of socket low head screws were loose and not torqued per assembly instruction causing the failure.The assignable root cause was determined to be due to a manufacturing issue.The issue related to manufacturing has been escalated to a non conformance.Device history record shows that the device was processed through the normal installation and inspection operations with no rework or nonconformities noted.The device met all required criteria at the time of release with no issues documented during the installation process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELYS SATELLITE STATION
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
DEPUY IRELAND
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17142917
MDR Text Key317313115
Report Number1818910-2023-12295
Device Sequence Number1
Product Code OLO
UDI-Device Identifier10603295519492
UDI-Public10603295519492
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K202769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4515-70-101
Device Catalogue Number451570101
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/30/2023
Initial Date FDA Received06/16/2023
Supplement Dates Manufacturer Received07/21/2023
Supplement Dates FDA Received07/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/31/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-