Model Number 4515-70-101 |
Device Problem
Positioning Failure (1158)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/30/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
|
|
Event Description
|
It was reported that the robotic assisted satellite station device, while being used with a robotic assisted base station device, the robot would not hold position.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was evaluated and the excessive movement on axis one of the robot was confirmed.The evaluation determined that the shaft axis of socket low head screws were loose and not torqued per assembly instruction causing the failure.The assignable root cause was determined to be due to a manufacturing issue.The issue related to manufacturing has been escalated to a non conformance.Device history record shows that the device was processed through the normal installation and inspection operations with no rework or nonconformities noted.The device met all required criteria at the time of release with no issues documented during the installation process.
|
|
Search Alerts/Recalls
|
|