The patient was undergoing a thrombectomy procedure in the m1 segment of the middle cerebral artery (mca) using a penumbra system ace 68 reperfusion catheter (ace68), a guidewire, a microcatheter, and a short sheath.It should be noted that the patient''s anatomy was tortuous.During the procedure, the physician experienced resistance while advancing the ace68 towards the target location under guidance of the microcatheter and the guidewire.The physician was unable to completely retract the ace68 and subsequently, the physician noticed on the angiography images that ace68 was fractured at the distal end.Therefore, the physician removed the proximal segment of the ace68 by hand and used a snare to remove the distal portion of the ace68 from the patient.The procedure was completed using a new ace68, the same short sheath, and the same microcatheter.There was no report of an adverse effect to the patient.
|
Evaluation of the first returned ace68 revealed stretching and a fracture on the distal end.This damage typically occurs due to retraction against resistance.Further evaluation revealed a kink near the fracture.This may indicate the device became pinned within patient anatomy.This may have contributed to resistance that resulted in the stretching and subsequent fracture of the device.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|