ROCHE DIAGNOSTICS COAGUCHEK XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Catalog Number 04625358019
Device Problem High Readings (2459)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/09/2023
Event Type  malfunction  
Manufacturer Narrative
E3-occupation is patient/consumer.The test strips were requested for investigation.The product has not been returned.If the product is returned in the future, a follow up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Product labeling states: "the coaguchek system uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas deviations can occur when compared to methods using other thromboplastins.However, those deviations between thromboplastins of different origin (e.G., rabbit based) are not specific to coaguchek products.Similar differences can be observed when a human recombinant thromboplastin based laboratory method is compared to other laboratory methods." the investigation did not identify a product problem.The cause of the event could not be determined.H3 other text : na.
 
Event Description
The customer complained of discrepant inr results with coaguchek xs meter serial number unknown compared to a laboratory using an unknown reagent.The results from the meter were 3.6 inr and 3.8 inr.The result from the laboratory at the same time the result was 2.2 inr.Therapeutic range was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key17145398
MDR Text Key318011932
Report Number1823260-2023-02009
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 06/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number04625358019
Device Lot NumberASKU
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/19/2023
Initial Date FDA Received06/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-