BAXTER HEALTHCARE CORPORATION BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number JC7790
Device Problems Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/16/2023
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter facility name: (b)(6).G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial the device was discarded and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a straight type blood set leaked.The dark blue slide clamp that is placed into the iv pump created a tear/partial sever in the iv tubing when a staff member unclamped it, causing the blood product to leak and the entire unit of blood to be contaminated.The issue was identified during an unspecified process step.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17164066
MDR Text Key317742847
Report Number1416980-2023-03028
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberJC7790
Device Lot NumberASKU
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/25/2023
Initial Date FDA Received06/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BLOOD
-
-