H4: the device was manufactured from november 03, 2022 - november 04, 2022.The actual device was not available; however, two (2) photographs of the sample were provided for evaluation.The photographs were visually inspected.The first photo shows a valve set in its packaging (pouch) and shows an oily residue/droplets on the inside of the pouch which appears to be on the device.The second photo is of the pouch with the valve set removed and the inside of the packaging has a clear substance, which is likely excess silicone oil on the inside of the pouch that possibly transferred from valve body to the inside of the packaging.The reported condition was verified.The cause of the condition was excess silicone used during the manufacturing assembly process, which was on the device and the inside of the pouch.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|