CALDERA MEDICAL BENESTA TRD Back to Search Results
Model Number CAL-TR1511
Device Problems Material Separation (1562); Material Deformation (2976); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2023
Event Type  malfunction  
Event Description
Trd outer cutter detached from device during the procedure.Second trd was used and procedure was completed.Caldera rep noticed the outer cutter was bent during the procedure, caused by torquing the device while inside the working channel of the hysteroscope.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENESTA TRD
Type of Device
BENESTA TRD
Manufacturer (Section D)
CALDERA MEDICAL
4360 park terrace drive
westlake village CA 91361
Manufacturer Contact
ryan troncoso
4360 park terrace drive
westlake village, CA 91361
MDR Report Key17167295
MDR Text Key318015163
Report Number3003990090-2023-01565
Device Sequence Number1
Product Code HIH
UDI-Device Identifier00857492008096
UDI-Public00857492008096
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberCAL-TR1511
Device Catalogue NumberCAL-TR1511
Device Lot NumberP01039
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/08/2023
Initial Date FDA Received06/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-