Model Number 862451
Device Problem Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/23/2023
Event Type  malfunction  
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Event Description
The customer reported that the monitor fell off its mounting arm while in use.There was no adverse event or patient harm reported.
 
Manufacturer Narrative
During onsite investigation by philips technicians and customer biomeds on (b)(6) 2023, it was determined that there is a push button on the monitor used to release the monitor from the connecting pins on the mounting arm.It appeared the button on the monitor was "sticking" and if pushed in, it would stick and not close back on the connecting pins.It was also noted that in this nicu setting, the arms are frequently being moved around infant cots/warmers in such a way that the button could be depressed while making those movements.The customer was provided replacements iv-mp70 mechasy quick release mount kits to resolve the issue.It has been concluded that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MP70
Type of Device
INTELLIVUE MP70
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key17167343
MDR Text Key317570458
Report Number9610816-2023-00291
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151681
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number862451
Device Catalogue Number862451
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/23/2023
Initial Date FDA Received06/20/2023
Supplement Dates Manufacturer Received07/06/2023
Supplement Dates FDA Received08/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/27/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-