Model Number 0673 |
Device Problems
Pacing Problem (1439); Failure to Sense (1559); Device Dislodged or Dislocated (2923); Positioning Problem (3009)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/02/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that this lead was unable to be successfully implanted as after being positioned it was moved, dropped and stuck out of the tissue for several times.The lead was repositioned multiple times and no signals were detected as there was no sensing and no pacing observed.Troubleshooting was performed by changing the cables and settings for lead testing, however, no changes were noted.The lead was then removed and replaced and the signals with the new lead were adequate and stable.No adverse patient effects were reported.The lead is expected to be returned.
|
|
Event Description
|
It was reported that this lead was unable to be successfully implanted as after being positioned it was moved, dropped and stuck out of the tissue for several times.The lead was repositioned multiple times and no signals were detected as there was no sensing and no pacing observed.Troubleshooting was performed by changing the cables and settings for lead testing, however, no changes were noted.The lead was then removed and replaced and the signals with the new lead were adequate and stable.No adverse patient effects were reported.The lead was returned for analysis.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance and inner insulation integrity.Measurements throughout these tests were within normal limits.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no anomalies.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observations of dislodged or dislocated, difficult to position, failure to sense and brady pacing not delivered when required.
|
|
Search Alerts/Recalls
|
|