Catalog Number EVO-FC-R-20-25-10-E
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Pneumonia (2011)
Event Type  Injury  
Event Description
Respiratory, thoracic, mediastinal - pneumonia; device related, procedure not related, pre-existing no; deficiency no.Respiratory, thoracic, mediastinal pneumonia - requiring medical intervention.Possible aspiration in context of stent insertion.Not definite.Resolved, treatment medical abc, no.
 
Manufacturer Narrative
Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
 
Manufacturer Narrative
Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
 
Event Description
Supplemental mdr being submitted to reflect changes to rpn/ product details updated from evo-fc-20-25-10-e to evo-fc-r-20-25-10-e.
 
Event Description
This supplemental report is being submitted due to completion of the investigation on the 02-nov-2023.
 
Manufacturer Narrative
This file was created from pmcf study to capture pneumonia.Device evaluation: the device evaluation could not be completed as the device or photographic evidence of the device was not returned for evaluation.With the information provided, a document-based investigation was conducted.Manufacturing records: prior to distribution, all evo-fc-r-20-25-10-e devices are subjected to functional checks and visual inspection to ensure device integrity.It may be noted that there has been discrepancies in rpn number provided.Review historical data: n/a.Instructions for use and/label: it should be noted that the instructions for use (ifu0067) which informs the user about the potential complications "those associated with upper gi endoscopy include, but are not limited to: allergic reaction to contrast or medication, aspiration, cardiac arrhythmia or arrest, fever, hemorrhage, hypotension, infection, perforation, reflux, respiratory depression or arrest, vomiting.As per medical advisor ¿ifu: aspiration; infection." there is no evidence to suggest the user did not follow the ifu or label.Image review: an image was not returned for evaluation.Root cause analysis: a definitive root cause could not be determined from the available information.A possible root cause could be attributed to patient condition related, as per instructions for use, aspiration and infection are listed as a potential complication.Confirmation of complaint: complaint is confirmed based on customer and/or rep testimony.Corrective action/correction: complaints of this nature will continue to be monitored for potential emerging trends.Summary of investigation: according to the customer, respiratory, thoracic, mediastinal - pneumonia.Confirmed quantity of 01 device, confirmed used.According to the initial reporter, resolved, treatment medical abc, no.As per medial advisors input "medical intervention s=4." investigation findings conclude that a possible root cause could be attributed to patient condition related, as per instructions for use, aspiration and infection are listed as a potential complication.Complaint is confirmed based on customer and/or rep testimony.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION® ESOPHAGEAL CONTROLLED-RELEASE STENT - FULLY COVERED
Type of Device
ESW PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer (Section G)
COOK IRELAND LTD
o halloran road
national technology park
limerick
Manufacturer Contact
sinead o'leary
o halloran road
national technology park
limerick 
MDR Report Key17169385
MDR Text Key317531460
Report Number3001845648-2023-00487
Device Sequence Number1
Product Code ESW
UDI-Device Identifier10827002551715
UDI-Public(01)10827002551715(17)190206(10)C1323153
Combination Product (y/n)N
PMA/PMN Number
K093619
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/06/2019
Device Catalogue NumberEVO-FC-R-20-25-10-E
Device Lot NumberC1323153
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location Hospital
Initial Date Manufacturer Received 04/11/2023
Initial Date FDA Received06/21/2023
Supplement Dates Manufacturer Received04/11/2023
04/11/2023
Supplement Dates FDA Received09/15/2023
12/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age91 DA
Patient SexMale
-
-