ALCON RESEARCH, LLC - ALCON PRECISION DEVICE CENTURION VISION SYSTEM, HANDPIECE TIP; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065752065
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that the system message that gas pressure was high was displayed and the ultrasound tip was bent during calibration.The surgery was completed after replacing the product with another one.There was no patient harm reported.
 
Manufacturer Narrative
One opened phacoemulsification tip in a wrench in a blister for the reported issue.The returned sample was visually inspected and was found to be nonconforming; it was bent at the cone to cannula transition.Wear was visible on the threads, back of flange and the nut corners, which is consistent with threading onto a handpiece.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The exact root cause could not be determined from the investigation performed.The returned sample is visually non-conforming with the phacoemulsification tip bent; therefore, a bent phacoemulsification tip was confirmed; however, how and when the phacoemulsification tip became bent could not be determined.Most likely root cause is handling during placement of the phacoemulsification tip onto the handpiece.The exact root cause for the bent phacoemulsification tip is unknown, therefore specific action with regards to this complaint cannot be taken.All phacoemulsification tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Any nonconformance, such as the phacoemulsification tip cannula bent exhibited on the returned opened sample, is removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, HANDPIECE TIP
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17169425
MDR Text Key317534924
Report Number2523835-2023-00302
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657520657
UDI-Public00380657520657
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065752065
Device Lot Number14XP3T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/26/2023
Initial Date FDA Received06/21/2023
Supplement Dates Manufacturer Received08/23/2023
Supplement Dates FDA Received09/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/28/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CENTURION SURGICAL PROCEDURE PAK
-
-