MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number AFAPRO28
Device Problems Material Integrity Problem (2978); Material Twisted/Bent (2981); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2023
Event Type  malfunction  
Event Description
It was reported that prior to a cryo ablation procedure during preparation of the catheter, an unknown system notice was received in dicating fluid was detected in the catheter/a leak was reconized and blocked the use of the console.The coaxial umbilical cable was replaced, the continue button pressed, and the console rebooted multiple times without resolution.The balloon catheter was replace d which resolved the issue. the case was completed with cryo. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the afapro28 balloon catheter with lot number 18451 was returned and analyzed.No anomalies were identified during external visual inspection of the balloon, shaft, and handle segments.The catheter smart chip data was reviewed.Data indicated the catheter was never used.No applications were performed.During functional testing, the console terminated the application and triggered system notice 50005 "the safety system detected fluid in the catheter and stopped the injection." pressure testing and inspection was performed on the sub-components of the balloon, handle, and shaft segments.During inspection and pressure testing of the shaft segment, a guide wire lumen kink and breach was observed 7.5 inches proximal to the catheter tip.During inspection of the shaft segment, a shaft breach was observed 7.5 inches from the catheter tip.In conclusion, the balloon catheter failed the returned product inspection due to the guide wire lumen breach and kink.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17170083
MDR Text Key317551741
Report Number3002648230-2023-00309
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeLO
PMA/PMN Number
P100010/S015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAFAPRO28
Device Catalogue NumberAFAPRO28
Device Lot Number18451
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/07/2023
Initial Date FDA Received06/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexMale
Patient Weight98 KG
-
-