Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/31/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the threads of the inserter were damaged during a procedure.There was no harm or impact to the patient.Attempts have been made and no further information is available.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of photos.Visual examination of the provided pictures identified the threads on the tip of the inserter to be damaged/deformed.Due to the quality of the picture it cannot be confirmed if broken/stripped.No further evaluation can be made from the provided picture.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|