Model Number URF-V2
Device Problems Break (1069); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned and evaluated, and the issue was confirmed.In addition to the bending tube being broken, and metal sticking out of the bending rubber, the additional evaluation findings are as follows: there was water invasion in the grip unit.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus that the uretero-reno videoscope had bending manipulation and insertion tube defects.There was no patient harm associated with the event.The device was returned and evaluated, and the bending tube was broken, and metal was sticking out of the bending rubber.This mdr (medical device report is being submitted to capture the reportable malfunction found during the device evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the bending section was damaged by application of physical stress during user handling.It was possible that damaged components of the bending section stuck into the bending section during user handling, leading to damage.However, the definitive root cause was unable to be determined.The event can be prevented by following the instructions for use which state: "ifu: urf-v2/v2r operation manual.Chapter 3 preparation and inspection: important information ¿ please read before use_ precautions:do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.Chapter 4 operation 4.1 warnings and cautions: operation do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.Do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17172151
MDR Text Key317629568
Report Number9610595-2023-09109
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170343582
UDI-Public04953170343582
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/31/2023
Initial Date FDA Received06/21/2023
Supplement Dates Manufacturer Received07/26/2023
Supplement Dates FDA Received08/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/20/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-