Model Number ATG80142 |
Device Problems
Break (1069); Deflation Problem (1149)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/26/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiry date: 12/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that during dilatation in the area of the brachio-cephalic vein leading into the superior vena cava, an area of very hard stenosis caused the balloon to be allegedly caught in the stenosis and the intima of the vein, whilst the inflation of the balloon caused the balloon to partially slip off of the balloon catheter.It was further reported that the balloon did not dislodge, but allegedly failed to deflate when negative pressure was applied via the inflation device.Reportedly, a guidewire was guided behind and another balloon was inflated behind the pta balloon in order to deflate it via external pressure.The catheter including all balloon material was removed successfully.There was no reported patient injury.
|
|
Manufacturer Narrative
|
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported break and deflation issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.D4 (expiry date: 12/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that during dilatation in the area of the brachio-cephalic vein leading into the superior vena cava, the inflation of the balloon allegedly caused the balloon to partially slip off of the balloon catheter.It was further reported that the balloon did not dislodge, but allegedly failed to deflate when negative pressure was applied via the inflation device.Reportedly, a guidewire was guided behind and another balloon was inflated behind the pta balloon in order to deflate it via external pressure.The catheter including all balloon material was removed successfully.There was no reported patient injury.
|
|