MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number ATG80142
Device Problems Break (1069); Deflation Problem (1149)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/26/2023
Event Type  malfunction  
Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiry date: 12/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during dilatation in the area of the brachio-cephalic vein leading into the superior vena cava, an area of very hard stenosis caused the balloon to be allegedly caught in the stenosis and the intima of the vein, whilst the inflation of the balloon caused the balloon to partially slip off of the balloon catheter.It was further reported that the balloon did not dislodge, but allegedly failed to deflate when negative pressure was applied via the inflation device.Reportedly, a guidewire was guided behind and another balloon was inflated behind the pta balloon in order to deflate it via external pressure.The catheter including all balloon material was removed successfully.There was no reported patient injury.
 
Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported break and deflation issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.D4 (expiry date: 12/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during dilatation in the area of the brachio-cephalic vein leading into the superior vena cava, the inflation of the balloon allegedly caused the balloon to partially slip off of the balloon catheter.It was further reported that the balloon did not dislodge, but allegedly failed to deflate when negative pressure was applied via the inflation device.Reportedly, a guidewire was guided behind and another balloon was inflated behind the pta balloon in order to deflate it via external pressure.The catheter including all balloon material was removed successfully.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD
Type of Device
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17180912
MDR Text Key317642975
Report Number2020394-2023-00442
Device Sequence Number1
Product Code DQY
UDI-Device Identifier00801741060731
UDI-Public(01)00801741060731
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K181323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberATG80142
Device Catalogue NumberATG80142
Device Lot Number93ZF0040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/27/2023
Initial Date FDA Received06/22/2023
Supplement Dates Manufacturer Received09/14/2023
Supplement Dates FDA Received09/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-