This report is based on information provided by a philips remote service engineer and bench engineer, and has been investigated by the philips complaint handling team.Philips received a complaint on the tempus pro patient monitor indicating that the device monitor would not boot to its main screen.The customer confirmed that another device was used to monitor the patient and there was no delay in providing the treatment, also the patient survived with no medical consequences reported.Remote support was provided, finding that the device won't boot beyond rdt splash screen.Monitor powers on but screen turns and remains black, device remains unresponsive, power button led continuously flashes green.The device was returned to philips for bench repair.The bench engineer confirmed the issue, the unit will not complete the boot cycle.Upon internal inspection the units trizeps board had been dislodged from the sbc.The trizeps was seated correctly, the device boots up and all device functions are performing per specifications.Device was soaked for 24 hours to ensure functionality has been restored.The unit was returned to the customer site.Based on the information available and results of additional analysis, no further action is necessary at this time.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.
|