SCHILLER AG TEMPUS LS; LOW ENERGY DEFIBRILLATOR Back to Search Results
Model Number 00-3020
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It has been reported to philips the device has internal discharge, charging not possible.Failed igbt self test while performing pm.
 
Manufacturer Narrative
The complaint was escalated for technical investigation and the results indicate that the error occurred because the connector pins were not fully and securely pressed in, to create a stable connection.It is therefore a production error on the part of estelec, the manufacturer of the dpm card.The customer was provided a replacement device to resolve the issue.It has been concluded that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS LS
Type of Device
LOW ENERGY DEFIBRILLATOR
Manufacturer (Section D)
SCHILLER AG
altgasse 68
baar 6341
SZ  6341
Manufacturer (Section G)
SCHILLER AG
altgasse 68
baar 6341
SZ   6341
Manufacturer Contact
jon yard
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key17183450
MDR Text Key317771487
Report Number3003832357-2023-00398
Device Sequence Number1
Product Code LDD
UDI-Device Identifier07613365002737
UDI-Public7613365002737
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number00-3020
Device Catalogue Number989706001681
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/01/2023
Initial Date FDA Received06/22/2023
Supplement Dates Manufacturer Received12/04/2023
Supplement Dates FDA Received02/01/2024
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-