• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. VISERA ELITE XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problem Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/25/2023
Event Type  malfunction  
Event Description
A customer reported no light from the scope when connected to the visera elite xenon light source.There were no reports of patient harm.
 
Manufacturer Narrative
The device was returned to an olympus repair facility and an evaluation of the device was performed.Upon inspection and testing of the unit, the reported problem of no light from the scope could not be reproduced.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
Correction to the previous follow-up report as h4 was incorrectly corrected.The device manufacturer date is 08oct2013 from the initial medwatch.
 
Manufacturer Narrative
Correction: correcting h4 - manufacturer date from oct 8, 2913 to feb.25, 2021 this report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to h4 of the initial medwatch a review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and because the phenomenon was not duplicated during device evaluation, the root cause of the phenomenon could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17187669
MDR Text Key318191804
Report Number3002808148-2023-06303
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170301995
UDI-Public04953170301995
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 08/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/25/2023
Initial Date FDA Received06/22/2023
Supplement Dates Manufacturer Received07/13/2023
08/21/2023
Supplement Dates FDA Received08/09/2023
08/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/08/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-