Model Number A22040A |
Device Problems
Material Discolored (1170); Scratched Material (3020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device was returned to olympus for evaluation.During inspection and testing, the allegation was confirmed.There was a crack in seal rubber (yellow).In addition to evaluation in b5, the seal rubber (gray) had discoloration.The guide screw was missing.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
|
Event Description
|
The customer reported to olympus, the inner sheath had a yellow packing crack on the hand side.There was no report of patient harm associated with this event.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was identified: scraping and discoloration on the inside of the tip beak.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The e1 "telephone number" field was corrected based on information available at the time of the initial submission.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been more than 17 years since the subject device was manufactured.Based on the results of the investigation, the insulation tip's damage was caused by wear and tear, improper handling and thermal and mechanical induced impact.The event can be prevented by following the instructions for use which state: 4 before use: "warning infection control risk properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing inspecting the product visually inspect the product.Make sure that it has: -- no corrosion -- no dents -- no scratches ceramic insulation at distal end visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning risk of injury impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product if the product is damaged or does not function properly, contact an olympus representative or an authorized service center." olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|