Model Number MCH-1000(I) |
Device Problem
No Flow (2991)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/27/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer notified cardioquip that the unit was not outputting any flow to the cardioplegia circuit when set to cold cpg.Cardioquip's investigation determined that two valves of the cpg valve assembly were nonfunctional and required replacement, causing the reported malfunction.Following replacement, the device passed inspection, and is fully operational.
|
|
Event Description
|
The customer has reported that the pump is not circulating as it should, with no alarms on the screen.After priming the device, the main pump is working.However, there is no flow in the cpg circuit when pressing cold.
|
|