Model Number 8886848813 |
Device Problem
Device Slipped (1584)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/05/2023 |
Event Type
malfunction
|
Event Description
|
According to the reporter, during the laparoscopic cholecystectomy, the surgeon used the clip applier to clip the cystic duct but failed.The clip clamped the cystic duct and then fell off.No device component fell into the cavity.A new clip reload was used to resolve the issue.There was no patient injury.
|
|
Manufacturer Narrative
|
Concomitant medical product: unk laprocl ins, unknown lapro-clip instrument, (lot #unknown) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but photos were available for evaluation.Visual inspection of the first photo noted a fully formed clip with the cartridge.The second photo noted a clip fully formed with one track not aligned and off to the side.It was reported that the clip did not hold tightly to the vessel and detached.The reported issue was confirmed.The most likely cause could not be established from the information available.The issue may occur if a side force or twist was applied to the clip track during application causing the clip track to not stay aligned.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|