Evaluation of the returned neuron max confirmed that the catheter was fractured, and the internal coil winds were unraveled at the hub.Evaluation revealed that the device was kinked at the id band and, upon inspection, was confirmed to be the location of the fracture.If the hub of the neuron max is torqued during manipulation of the device within the reported patient¿s tortuous anatomy, damage such as a kink and subsequent fracture may occur.Further evaluation revealed a kink and ovalization on the device.This damage was likely incidental to the reported complaint and likely occurred during packaging for return to penumbra.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a medical procedure using a neuron max 6f 088 long sheath (neuron max).It was noted that the patient¿s anatomy was tortuous.During the procedure, while injecting contrast through the neuron max, the physician noticed the neuron max had unraveled at the hub.Therefore, the neuron max was removed.The procedure was completed using a new neuron max.There was no report of an adverse effect to the patient.
|