LUTONIX, INC LUTONIX 018 DRUG COATED DILATATION CATHETER; DRUG COATED BALLOON PTA CATHETER
|
Back to Search Results |
|
Model Number 9014 |
Device Problems
Difficult to Insert (1316); Material Puncture/Hole (1504); Failure to Advance (2524); Device-Device Incompatibility (2919)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/30/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during an angioplasty procedure in the superficial femoral artery, the guidewire was allegedly failed to go through the wire lumen.It was further reported that the guidewire allegedly protruded through the either side of the nylon material on the balloon or the peel away protector.The procedure was completed by using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiration date: 01/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Device not returned.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported guidewire advancement failure and material puncture could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 01/2026), g3 h11: h6 (device) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that during an angioplasty procedure in the superficial femoral artery, the guidewire was allegedly failed to go through the wire lumen.It was further reported that the guidewire allegedly protruded through the either side of the nylon material on the balloon or the peel away protector.The procedure was completed by using another device.There was no reported patient injury.
|
|