• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HALOGEN LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HALOGEN LIGHT SOURCE Back to Search Results
Model Number CLK-4
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/29/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation, and the customer's allegation was confirmed; due to a faulty transformer unit.The additional evaluation findings are as follows: the washer under the screw on rear panel was missing, and the lamp harness was deformed.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The olympus field service engineer reported (on behalf of the customer) that the halogen light source lamp does not light.The issue was found during maintenance.There were no reports of patient harm and no procedural involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to a faulty transformer unit.However, the definitive root cause was unable to be determined.The lamp was found to be a third party lamp.Olympus always recommends to use olympus md-151 halogen lamp.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Type of Device
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17201167
MDR Text Key318475910
Report Number3002808148-2023-06380
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170024528
UDI-Public04953170024528
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K993041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 08/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLK-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/29/2023
Initial Date FDA Received06/26/2023
Supplement Dates Manufacturer Received07/18/2023
Supplement Dates FDA Received08/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/24/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-