MAUDE Adverse Event Report: ENCORE MEDICAL L.P EMPOWR ACET DM, HEAD TRIAL, THREADED OFFSET SLEEVE, -3.5MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P EMPOWR ACET DM, HEAD TRIAL, THREADED OFFSET SLEEVE, -3.5MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 803-15-111
Device Problem Material Integrity Problem (2978)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 05/30/2023
Event Type  malfunction  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event Description
Instrument failure - 1 hour delay in surgery.
 
Manufacturer Narrative
The agent reported (surgeon had difficulty dislocating trial hip configuration.So, he used a bone saw and chisel to cut apart the head and sleeve trials.Rendering them completely useless for future cases.Female, 52).This investigation is limited in scope as only partial information was provided to djo surgical - austin for review.The revised item(s) was not returned for examination and the item and or lot number(s) was not provided.To adequately investigate this event, the part and lot number(s) are necessary.If this information is submitted at a future date, this investigation will be re-evaluated.Customer complaint history of the reported item(s) showed no present trends or on-going issues that are needing a review.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The surgeon performed this revision to remedy the patient's condition.This complaint will be closed pending receipt of additional information.No further action is deemed necessary at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR ACET DM, HEAD TRIAL, THREADED OFFSET SLEEVE, -3.5MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758
MDR Report Key17205817
MDR Text Key317913856
Report Number1644408-2023-00787
Device Sequence Number1
Product Code LPH
UDI-Device Identifier00190446689775
UDI-Public00190446689775
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203557
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number803-15-111
Device Catalogue Number803-15-111
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/30/2023
Initial Date FDA Received06/26/2023
Supplement Dates Manufacturer Received08/14/2023
Supplement Dates FDA Received09/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
803-15-140.
Patient Outcome(s) Other;
Patient Age52 YR
Patient SexFemale
-
-