Model Number L301 |
Device Problems
Over-Sensing (1438); Pacing Problem (1439); Incorrect, Inadequate or Imprecise Result or Readings (1535); Inappropriate or Unexpected Reset (2959)
|
Patient Problems
Loss of consciousness (2418); Syncope/Fainting (4411)
|
Event Date 06/13/2023 |
Event Type
Injury
|
Event Description
|
It was reported that this pacemaker had entered safety mode.The patient was pacer dependent and reportedly passed out three times due to unipolar oversensing and pacing inhibition while the pacemaker was in safety mode.Subsequently, this pacemaker was explanted, replaced, and returned for analysis.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
The product has been received for analysis.This report will be updated upon completion of analysis.
|
|
Event Description
|
It was reported that this pacemaker had entered safety mode.The patient was pacer dependent and reportedly passed out three times due to unipolar oversensing and pacing inhibition while the pacemaker was in safety mode.Subsequently, this pacemaker was explanted, replaced, and returned for analysis.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode and that bradycardia therapy remained available.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance put this device at risk of experiencing transient voltage decreases (and associated resets) during periods of high-power consumption.When battery voltage drops below a minimum threshold, a system reset is performed.If three system resets occur within a 48-hour period, the device is designed to immediately enter safety mode operation to maintain back-up pacing with pre-defined, non-programmable settings.
|
|