Model Number WA22702S
Device Problems Therapeutic or Diagnostic Output Failure (3023); Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2023
Event Type  malfunction  
Event Description
Olympus bipolar machine not functioning for procedure.New maching obtained, still issue.Obtained new plasma loop and it worked.Appears plasma loop the issue.No injury and no significant delays.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OLYMPUS
Type of Device
ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL
Manufacturer (Section D)
OLYMPUS CORPORATION OF THE AMERICAS
3500 corporate pkwy.
center valley PA 18034
MDR Report Key17216215
MDR Text Key318026964
Report Number17216215
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761085325
UDI-Public(01)14042761085325(11)230125(17)251210(10)1000110122
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 06/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA22702S
Device Catalogue NumberWA22702S
Device Lot Number1000110122
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA06/14/2023
Date Report to Manufacturer06/28/2023
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/28/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age26280 DA
-
-