The present problem with creatinine testing after installation of new bun/creatinine sensor cartridges has occurred on a regular basis since we acquired the nova stat profile prime plus instruments, and has occurred on both instruments.The problem always occurs after installation of a new bun/creatinine cartridge, however, bias trends for patient results vary; patient results may be 50% above or below expected concentrations.The problem has occurred among multiple lots and is not lot-specific; within one lot there may be only a single cartridge that produces unacceptable patient results upon installation.The occurrence seems to be random in frequency (1-3 total occurrences per year).In all instances, qc does not indicate unacceptable instrument performance.Identification of creatinine problems only occurs through evaluation of patient specimens.
|