Model Number CLV-190
Device Problems Fluid/Blood Leak (1250); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/23/2023
Event Type  malfunction  
Event Description
Customer returned the demo device reported with cosmetic appearance found in preparation for use.There was no patient involvement on this reported issue.Device return evaluation found liquid backflow into s socket (connector outer tube).This report is being submitted due to the finding of liquid backflow into s socket (liquid intrusion into the tube) identified during device evaluation.
 
Manufacturer Narrative
The subject device was evaluated.Device evaluation found poor protection bar fixation in s socket and as stated in section b5, liquid backflow into the s socket was observed.Additionally, inspection found scratch on the top cover exterior.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event is caused by liquid backflow into the s socket.A root cause could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17219359
MDR Text Key318306254
Report Number3002808148-2023-06502
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/02/2023
Initial Date FDA Received06/28/2023
Supplement Dates Manufacturer Received07/12/2023
Supplement Dates FDA Received08/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/05/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-