• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN SURGICAL, S.A. DAFILON BLUE 3/0 (2) 75CM DSMP19 (M); OTHER SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN SURGICAL, S.A. DAFILON BLUE 3/0 (2) 75CM DSMP19 (M); OTHER SUTURE Back to Search Results
Model Number C0936740
Device Problems Dull, Blunt (2407); Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/19/2023
Event Type  malfunction  
Event Description
It was reported an issue with dafilon suture.The client reported that the needle does not pierce.No patient injury reported, no consequences.Additional information has been requested.
 
Manufacturer Narrative
If additional information becomes available a follow up report will be submitted.
 
Manufacturer Narrative
Analysis and results: there are no previous complaints of this code-batch.We manufactured and distributed in the market (b)(4) units of this code-batch.There are no units in our stock.We have not received any sample for analysis.Without any closed and/or defective sample we cannot carry out an analysis in order to take a decision.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil usp/ep and b.Braun surgical requirements.As stated in the instructions for use of the product, care should be taken to avoid damaging the needle when using the suture material.Always grasp the needle in a section 1/3 to ½ of the distance from the fiber attachment end to the needle point, never at the end where the fibre is attached or the needle point.Grasping the needle at the area of its point could impair the penetration performance and cause a fracture of the needle.Grasping the needle close to the fibre attachment end could cause bending, breakage.Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done, and the case is not confirmed due to lack of evidence.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyse it.Please note that when no samples are received our analysis is very limited.We regret any inconvenience this issue may have caused and thank you for your collaboration.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAFILON BLUE 3/0 (2) 75CM DSMP19 (M)
Type of Device
OTHER SUTURE
Manufacturer (Section D)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP  08191
Manufacturer (Section G)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
Manufacturer Contact
martina laporte
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
MDR Report Key17224192
MDR Text Key318087297
Report Number3003639970-2023-00228
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K151165
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC0936740
Device Catalogue NumberC0936740
Device Lot Number622504
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/22/2023
Initial Date FDA Received06/29/2023
Supplement Dates Manufacturer Received07/26/2023
Supplement Dates FDA Received07/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-