Analysis and results: there are no previous complaints of this code-batch.We manufactured and distributed in the market (b)(4) units of this code-batch.There are no units in our stock.We have not received any sample for analysis.Without any closed and/or defective sample we cannot carry out an analysis in order to take a decision.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil usp/ep and b.Braun surgical requirements.As stated in the instructions for use of the product, care should be taken to avoid damaging the needle when using the suture material.Always grasp the needle in a section 1/3 to ½ of the distance from the fiber attachment end to the needle point, never at the end where the fibre is attached or the needle point.Grasping the needle at the area of its point could impair the penetration performance and cause a fracture of the needle.Grasping the needle close to the fibre attachment end could cause bending, breakage.Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done, and the case is not confirmed due to lack of evidence.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyse it.Please note that when no samples are received our analysis is very limited.We regret any inconvenience this issue may have caused and thank you for your collaboration.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
|