Model Number 25-60-70
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/30/2023
Event Type  malfunction  
Manufacturer Narrative
A.1.-a.5.Patient information were not provided.H10: livanova deutschland manufactures the flow sensor measurement.The incident occurred in united states.A livanova field service engeneet was dispatched to the customer and found the flow sensor and the flow module were out of tolerance.They have been replaced.Performed flow test performance passed.Performed all functional checks per tsi system is functional and ready for use.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland has received a report that, during a procedure, patient blood flow will not correlate when the rpm maxs out.There was no patient injury.
 
Manufacturer Narrative
H10: complaints database analysis revealed no similar event since unit installation in 2019.It cannot be ruled out that the root cause of the reported event was a defective flow sensor and module due to wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CP5 FLOW MODULE
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17225143
MDR Text Key318224288
Report Number9611109-2023-00293
Device Sequence Number1
Product Code DWA
UDI-Device Identifier04033817900733
UDI-Public(01)04033817900733(11)190401
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25-60-70
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/30/2023
Initial Date FDA Received06/29/2023
Supplement Dates Manufacturer Received07/28/2023
Supplement Dates FDA Received08/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-