Model Number 25-60-70 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/30/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A.1.-a.5.Patient information were not provided.H10: livanova deutschland manufactures the flow sensor measurement.The incident occurred in united states.A livanova field service engeneet was dispatched to the customer and found the flow sensor and the flow module were out of tolerance.They have been replaced.Performed flow test performance passed.Performed all functional checks per tsi system is functional and ready for use.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland has received a report that, during a procedure, patient blood flow will not correlate when the rpm maxs out.There was no patient injury.
|
|
Manufacturer Narrative
|
H10: complaints database analysis revealed no similar event since unit installation in 2019.It cannot be ruled out that the root cause of the reported event was a defective flow sensor and module due to wear.
|
|