MAUDE Adverse Event Report: DIASORIN INC. 2 TYPE SPECIFIC ANTIBODY TEST; ENZYME LINKED IMMUNOSORBENT ASSAY, HERPES SIMPLEX VIRUS, HSV-2

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DIASORIN INC. 2 TYPE SPECIFIC ANTIBODY TEST; ENZYME LINKED IMMUNOSORBENT ASSAY, HERPES SIMPLEX VIRUS, HSV-2 Back to Search Results
Device Problem False Positive Result (1227)
Patient Problem Insufficient Information (4580)
Event Date 05/01/2023
Event Type  Injury  
Event Description
Patient had a hsv 2 positive test on the diasorin hsv 1 and 2 igg test, (index value = 1.22), followed up with a western blot which was negative for hsv 1 and 2.Date given is the western blot confirmatory test.Reference report: mw5118937.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2 TYPE SPECIFIC ANTIBODY TEST
Type of Device
ENZYME LINKED IMMUNOSORBENT ASSAY, HERPES SIMPLEX VIRUS, HSV-2
Manufacturer (Section D)
DIASORIN INC.
MDR Report Key17226014
MDR Text Key318242104
Report NumberMW5118938
Device Sequence Number1
Product Code MYF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date 06/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/28/2023
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age37 YR
Patient SexMale
Patient Weight73 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-